| | DESCRIPTION: POWERTRONIX Medical Isolation Transformers are recognized by UL 2601 and built to meet the low leakage current requirements in today's medical equipment. Automatic resetable thermal fuses are included, providing protection from overload and short circuit conditions.
APPLICATIONS: Medical Isolation Transformers are used in medical and dental instruments, EEG monitors, electro- surgical equipment, medical lasers, NMR equipment, scanners, and ultrasonic devices. We also provide Medical Power Supplies. | |
| SPECIFICATIONS: | | | . Power: | 100 VA - 5000 VA | | . Frequency: | 47 Hz - 63 Hz | | . Primary: | 2 x (0-100-120) Volt | | . Secondary: | 2 x 120 Volt | | . Isolation: | 4,000 VRMS | | . Creepage: | Minimum 8 mm | | . Leakage current: | Less than 50 uA | | . Voltage tolerance: | Better than 3% | | | | SAFETY APPROVALS: | | | . UL 2601, UL 544 | (File # E 151895) | | . UL 506 | (File # E 142730) | | | | | BUILT TO: | | | . IEC 601-1, EN 60601.1 | | | . VDE 750 Part 1 | | | |
|